• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 14F; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 14F; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 000714
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that allegedly the feeding tube unintentionally dislodged from the patient.It was reported that the balloon ruptured and leaked.There was no reported patient injury.
 
Additional Manufacturer Narrative
Complete manufacturing review could not be conducted for the investigation as the lot number is unknown.One 14 fr tri-funnel repl gastrostomy feeding tube was returned for evaluation.Functional, tactile, visual, microscopic, and measurement evaluations were performed.The investigation is confirmed for the reported deflation, as a 3 mm partially circumferential crack at the proximal end of the membrane attachment was found.Leaking occurred through the crack when the inflation lumen was infused.Magnification showed evidence of tearing and a relatively rough sharp edged granular surface at this crack.The definitive root cause could not be determined based upon available information.It is unknown if procedural issues contributed to the reported event; the damage to the balloon evidently occurred during use but the cause is unknown.A review of baw0732741, rev.2 was performed, however, the cause of this event is unknown and therefore the appropriate section in the ifu, if any, is uncertain.If manufacturing related, a review of product labeling documents is not applicable.Product labeling, such as indications, warnings, precautions, cautions, possible complications, and contraindications, would not have prevented this issue.
 
Event or Problem Description
It was reported that allegedly the feeding tube unintentionally dislodged from the patient.It was reported that the balloon ruptured and leaked.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 14F
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7768048
Report Number3006260740-2018-01997
Device Sequence Number19345428
Product Code KNT
UDI-Device Identifier00801741037122
UDI-Public(01)00801741037122
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000714
Device Catalogue Number000714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer11/19/2018
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-