No medical records or medical images have been made available to the manufacturer.As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Complete manufacturing review could not be conducted for the investigation as the lot number is unknown.One 14 fr tri-funnel repl gastrostomy feeding tube was returned for evaluation.Functional, tactile, visual, microscopic, and measurement evaluations were performed.The investigation is confirmed for the reported deflation, as a 3 mm partially circumferential crack at the proximal end of the membrane attachment was found.Leaking occurred through the crack when the inflation lumen was infused.Magnification showed evidence of tearing and a relatively rough sharp edged granular surface at this crack.The definitive root cause could not be determined based upon available information.It is unknown if procedural issues contributed to the reported event; the damage to the balloon evidently occurred during use but the cause is unknown.A review of baw0732741, rev.2 was performed, however, the cause of this event is unknown and therefore the appropriate section in the ifu, if any, is uncertain.If manufacturing related, a review of product labeling documents is not applicable.Product labeling, such as indications, warnings, precautions, cautions, possible complications, and contraindications, would not have prevented this issue.
|