VASCUTEK LTD GELWEAVE; GELWEAVE VALSALVA Back to Search Results
Catalog Number 730032ADP
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes Cardiac Arrest (1762); Infarction, Cerebral (1771); Thrombus (2101); Sudden Cardiac Death (2510)
Date of Event 07/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).On (b)(6) postoperatively, 6 hours after the scheduled surgery, the patient had a massive infarction complicated by cardiac arrest that required a coronography showing left coronary artery occlusion warranting emergency resurgence for bypass surgery.Just before re-intervention, the patient had a cardiac arrest requiring ecls.During reintervention, thrombus of the aortic pellet was found to mold the valve and aortic prosthesis and to obstruct the common trunk.Thrombus extraction and double bypass were performed.During follow-up, the patient presented a multi-organ failure leading to his death on (b)(6) 2018 in intensive care at day 1.(b)(4).The manufacturing and qc records for this batch were retrieved and reviewed.Batch was manufactured to design specification.No further complaints have been received from this batch to date.Similar events review that was carried out for gelweave valsalva grafts with occlusion/thrombosis events gave a low occurrence rate of (b)(4).Vascutek are awaiting further information from site.
 
Event or Problem Description
Event was reported to vascutek as follows: gelweave valsalva graft was used for bentall procedure.Extensive thrombosis of the vascular graft occurred few hours after surgery, requiring surgical revascularization.Patient died of massive infarction.
 
Event or Problem Description
This report is being submitted as follow up # 1 for mfr.Report to provide additional information that was received from the customer.More detailed event description was received on 19 jul 18: the patient was hospitalised for bentall elective surgery for annulo-ectasia disease with severe aortic leakage.Valved tube made of valsalva 32 tube and edwards carpenter perimount bio prosthesis of 27 gauge.Simple intervention clamping 74 min, cec 85 min, total duration 190 min.The immediate follow-up was simple correct hemodynamics, (no transfusion, no haemostatic products).On 6 july postoperatively, 6 hours after the scheduled surgery, the patient had a massive infarction complicated by cardiac arrest that required a coronography showing left coronary artery occlusion that lead to the emergency bypass surgery.Just before re-intervention, the patient had a cardiac arrest requiring ecls.During reintervention, thrombus of the aortic pellet was found to mold the valve and aortic prosthesis and to obstruct the common trunk.Thrombus extraction and double bypass were performed.During follow-up, the patient presented a multi-organ failure leading to his death on (b)(6) 2018 in intensive care at day 1.
 
Additional Manufacturer Narrative
Conclusion: 4315 cause not established.Manufacturer could not carry out further device analysis as it remains in situ.However, in the event intake form provided by the customer it was noted that the event was not caused by thrombopenia induced by heparin.Anapath evaluation report of thrombus showed following results: a circular thrombus 3.5 cm in diameter and 1.5 to 2 cm in height with a small "outgrowth" perpendicular to the center of 1.5 cm in height.A detached fragment of 1 x 0.2 cm.Cut transversally and included in total in 5 blocks (blocks 1 to 5).On the different plans examined, a fibrino-hemorrhagic material containing blood figured elements was observed, with quite a few polynuclear.A beginning of organization in streaks was observed.Focally, on two blocks (2 and 4) the presence of exogenous material, bluish, thrombus molded were observed, the nature of which cannot be specified.Vascutek ltd now considers this complaint closed.Further action is not planned; however, the issue will be tracked and trended as part of the on-going complaints trending and reporting process and if an adverse trend develops action may be taken at that time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GELWEAVE
Common Device Name
GELWEAVE VALSALVA
Manufacturer (Section D)
VASCUTEK LTD
newmains avenue
inchinnan business park
renfrewshire, PA4 9 RR
UK  PA4 9RR
MDR Report Key7768055
Report Number9612515-2018-00010
Device Sequence Number2050465
Product Code DSY
UDI-Device Identifier05037881120140
UDI-Public05037881120140
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K013022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/07/2018
Report Date (Section F) 07/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Catalogue Number730032ADP
Device Lot Number17128740-0857
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/11/2018
Device Age6 MO
Event Location Hospital
Date Report to Manufacturer07/12/2018
Initial Date Received by Manufacturer 07/12/2018
Supplement Date Received by Manufacturer07/12/2018
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Required Intervention;
Patient Age75 YR
Patient Weight87
-
-