• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE Back to Search Results
Catalog Number CAT8TORQ85
Medical Device Problem Codes Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the cat8 was ovalized approximately 81.5 though 86.0 cm from the hub.Conclusions: evaluation of the returned cat8 revealed that the distal tip was ovalized.If the peel-able sheath supplied by penumbra is not used to protect the distal tip of the catheter during insertion, damage such as ovalization may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure using indigo system aspiration catheter 8 (cat8).During the procedure, the physician experienced resistance while attempting to re-insert the cat8 into a non-penumbra sheath and subsequently, the distal tip of the cat8 became kinked; therefore, it was removed.The procedure was completed using the same sheath and a new cat8.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 8
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7768197
Report Number3005168196-2018-01547
Device Sequence Number11278013
Product Code DXE
UDI-Device Identifier00814548016252
UDI-Public00814548016252
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2021
Device Catalogue NumberCAT8TORQ85
Device Lot NumberF83254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2018
Initial Report FDA Received Date08/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age21 YR
Date Report Sent to FDA01/01/2005
-
-