| Model Number |
MMT-523NAH |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that the insulin pump received a software error alarm.Blood glucose was unknown at time of incident.Troubleshooting was completed.Customer was advised to discontinue use of the insulin pump and revert to a back-up plan per healthcare provider's instruction.The customer will receive a replacement insulin pump and will return the one they currently use for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed displacement test, rewind test, basic occlusion test, prime test, excessive no delivery test, occlusion test, self test and unexpected restart error test.Insulin pump passed all operating currents tested within the spec range and passed off no power test.Insulin pump was tested with no software error bad pointer alarm noted.Insulin pump received with cracked reservoir tube cracked reservoir tube window, cracked case near display window corners and cracked on display window noted.(b)(4).
|
| |
|