• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. MOTOR LEGEND EHS STYLUS REFURB; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. MOTOR LEGEND EHS STYLUS REFURB; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number EM200-R
Medical Device Problem Codes Overheating of Device (1437); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report inconclusive.The device has been returned and is still pending their evaluation results.Initial test was conducted and results showed that device overheated above threshold.Further evaluation is being performed to verify the cause and/or contributing factor of the overheating.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with no reported malfunction.It was reported that there was no patient or staff impact.Repair escalated to product event due to information received with the device indicating that the device overheated.
 
Additional Manufacturer Narrative
Report confirmed.Initial temperature tests were conducted and results showed that device overheated above threshold at 8.1 deg f/sec.Further evaluation determined there were no signs of mechanical wear and tear on the components and test data indicated that all electrical components were operating as expected.The likely cause of the device overheating could not be determined.There is no evidence that any other failures during analysis were found that may have contributed to this event.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR LEGEND EHS STYLUS REFURB
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7768559
Report Number1045254-2018-00359
Device Sequence Number11354193
Product Code HBE
UDI-Device Identifier00613994422002
UDI-Public00613994422002
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K012456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM200-R
Device Catalogue NumberEM200-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Initial Date Received by Manufacturer 07/19/2018
Supplement Date Received by Manufacturer07/19/2018
08/08/2018
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date08/17/2018
10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-