| Model Number |
ZXR00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
02/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device/lot history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tecnis symfony multifocal intraocular lens (model zxr00 +12.5 diopter) was explanted from patient¿s left eye in secondary procedure.Reportedly, patient was experiencing a hazy film with their vision, like looking through soapy glass.During the explant procedure an incision enlargement was performed; however, there was no patient injury, no vitrectomy and no suture.A lens model aab00 +12.5 diopter was used as replacement for the patient.Patient is doing good post-operative with no pain or discomfort.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on: 8/30/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The product is inspected by a qualified inspector using a microscope with a 12x magnification, no anomalies were identified.The complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|