| Model Number |
030453 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, while cutting the pulmonary vessel in a laparoscopic video-assisted thoracoscopic lobectomy, the surgeon found that the handle could not be squeezed and it could not fire the reload.The green button was pressed.They used another device to continue with the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.The visual inspection noted that the loading unit was pre-fired and engaged in interlock with the jaws open.Functional testing of the loading unit found no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the loading unit from firing a second time by ceasing the placement of staples and tissue transection and prevent patient harm.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|