• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926008220
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr: stent delivery system was returned for analysis.A visual examination of the stent found that the first proximal stent row was damaged with stent struts lifted on one side.Distal stent rows 1-2 were damaged with stent struts lifted and pulled distally.The undamaged crimped stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 18-jul-2018.It was reported that crossing difficulties were encountered.A 2.25x8mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed stent damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7769611
Report Number2134265-2018-07051
Device Sequence Number2217800
Product Code NIQ
UDI-Device Identifier08714729839965
UDI-Public08714729839965
Combination Product (Y/N)N
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2019
Device Model NumberH7493926008220
Device Catalogue Number39260-0822
Device Lot Number21864856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2018
Initial Report FDA Received Date08/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-