Model Number H965430600
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a wallstent biliary uncovered stent was to be used to treat a biliary obstruction during a stent placement procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the stent migrated immediately after deployment.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Boston scientific has been unable to obtain additional information regarding the circumstances surrounding this event to date.Should additional relevant details become available a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Problem code 3009 captures the reportable event of stent positioning issue.Investigation results: a fully deployed wallstent biliary stent and delivery system were received for analysis.Visual examination of the returned device found uncrossed and kinked stent wires at one end of the stent.The stent was measured to be within specifications.No other issues were noted to the stent and delivery system.The damages noted to the stent were most likely a result of unintentional use error during post procedure handling or re-packing of the device.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.Stent misplacement with a failure effect of stent migration is noted within the dfu as a potential adverse events associated with the use of this device.Therefore, the most probable root cause is known inherent risk of device.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
 
Event or Problem Description
It was reported to boston scientific corporation on july 21, 2018 that a wallstent biliary uncovered stent was to be used to treat a biliary obstruction during a stent placement procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the stent migrated immediately after deployment.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Boston scientific has been unable to obtain additional information regarding the circumstances surrounding this event to date.Should additional relevant details become available a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key7770209
Report Number3005099803-2018-02517
Device Sequence Number1758111
Product Code FGE
UDI-Device Identifier08714729204688
UDI-Public08714729204688
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K982184
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2020
Device Model NumberH965430600
Device Catalogue Number43060
Device Lot Number0021929701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Initial Date Received by Manufacturer 07/21/2018
Supplement Date Received by Manufacturer01/24/2019
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date02/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-