BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ7578
Medical Device Problem Codes Peeled/Delaminated (1454); Unraveled Material (1664)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that after an angioplasty procedure in the cephalic vein, unraveled material was found on the tip of the balloon, upon removal from the patient.No further treatment was needed.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found the outer layer to be peeled from the balloon throughout the length of the balloon.No unraveled fibers or fiber disturbance was noted to the balloon.Therefore, the investigation is confirmed for peeled outer layer.The investigation is unconfirmed for the reported unraveled material.The definitive root cause for the identified peeling outer layer or reported unraveled material could not be determined based upon the available information.An investigation is currently open to address the peeling outer layer issue.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that after an angioplasty procedure in the cephalic vein, unraveled material was found on the tip of the balloon, upon removal from the patient.No further treatment was needed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7770530
Report Number2020394-2018-01412
Device Sequence Number1909687
Product Code DQY
UDI-Device Identifier00801741063015
UDI-Public(01)00801741063015
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model NumberCQ7578
Device Catalogue NumberCQ7578
Device Lot NumberRECN0250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2018
Initial Date Received by Manufacturer 07/13/2018
Supplement Date Received by Manufacturer10/11/2018
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-