No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported that allegedly the hub of the handle allegedly broke during an attempt to deploy the stent graft.Reportedly, the endovascular stent graft delivery system was removed without incident, and another manufacturer's device was used instead to complete the procedure.There was no reported patient injury.
|
|
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Investigation summary: based on the investigation of the returned catheter sample it was confirmed that the user could only partially deploy the stent graft.The outer catheter was found detached from the grip which made a complete deployment impossible.Therefore, as a result of the investigation performed the reported issue was confirmed.However, based on the available information and the evaluation of the returned sample, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product it was found that the instructions for use (ifu) sufficiently address the potential risks.Regarding preparation of the device the ifu states that 'prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.' regarding the anatomy of the placement site the ifu states: 'the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure' and 'do not kink the delivery catheter or use excessive force during delivery to the target lesion.' furthermore, the ifu states: 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.' h10: d4 (expiry date: 01/2021), g4 h11: h3, h6 (device code + description2), (eval code & desc - results 1 & conclusion 1) h11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported that allegedly the hub of the handle allegedly broke during an attempt to deploy the stent graft.Reportedly, the endovascular stent graft delivery system was removed without incident, and another manufacturer's device was used instead to complete the procedure.There was no reported patient injury.
|