The associated complaint devices were returned and evaluated.Devices (b)(4).Were not returned for evaluation.The lab analysis for the femoral, right tibial articulating insert, and left tibial articulating insert scratches on the femoral condyle articular surfaces that are along the radius are likely the result of articulation of the knee.The source of other scratches could not be determined from the information provided.No material or manufacturing deviations were observed in the course of this investigation.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch/failure mode.The clinical/medical team concluded, this complaint was re-opened because the product returned.However, the only clinical information that has been provided is three pages of chart sticks that have been reviewed.No additional supporting documents have been provided to determine a root cause for the reported issues or patient impact.Therefore, no further clinical/medical assessment is warranted at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.
|