| Model Number |
H700123 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During the procedure with the patient on the table, the system was turned on without success.The system was rebooted which did not resolve the issue.The procedure was cancelled and will be rescheduled.There were no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided to abbott and the investigation performed, changes made to the system ini (initialization) configuration settings prevented a normal boot sequence or initialization.The configuration settings within the initialization file were corrected to resolve the issue.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.It is unknown how or when the configuration settings were changed.
|
| |
|