Non-numerical prothrombin time (pt) and pt international normalized ratio (inr) results, flagged with "no coagulation", were obtained on a patient sample on a sysmex cs-2100i system with the thromborel s reagent.The sample was repeated on the same system, resulting in discordant, falsely elevated pt and pt inr results.The discordant pt inr result was auto-validated and reported to the physician(s), who questioned the result.The sample was re-centrifuged and repeated on the same system, resulting lower.The patient's blood was also redrawn and run on the same system, resulting lower.The pt inr result, obtained on the redrawn sample, was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated pt and pt inr results.
|