• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE INNOVA¿; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE INNOVA¿; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number H74939293051530
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event or Problem Description
Same case as mdr id: 2134265-2018-07263.It was reported that partial stent deployment occurred.The target lesion was located in the heavily calcified popliteal artery.A 5 x 150 x 130 innova¿ was selected for use and deployment attempted.Only a partial amount of the stent deployed from the delivery system.The delivery system with stent was removed from the patient.The physician then chose another 5 x 150 x 130 innova¿ and deployment was attempted; however, the same thing occurred.The procedure was completed with angioplasty only.There were no patient complications and the patient condition post-procedure was fine.
 
Additional Manufacturer Narrative
Describe event or problem: updated.(b)(4).
 
Event or Problem Description
It was further reported that the target lesion was over 90% stenosed and not too severely calcified.Vascular access was obtained via a contralateral approach.The lesion was pre-dilated.The first 5 x 150 x 130 innova¿ was advanced over a.014 guidewire.The second 5 x 150 x 130 innova¿ was advanced over a.035 guidewire.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the outer shaft, middle shaft, inner sheath and the remainder of the device were checked for damage.There was a kink at the nosecone.The stent was partially deployed approximately 4.3cm from the distal end of the markerband.The thumbwheel was rotated and the stent deployed as designed.The device was opened and inspected for any internal damage.No damage was noticed.Per the returned information the physician felt resistance and stopped the case and pulled the device from the patient which partially deployed the stent farther than normal.Inspection of the remainder of the device, revealed no other damage or irregularities.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.The innova dfu states that a stiff 0.035 in guidewire is strongly recommended for deployment of the stent, especially for tortuous anatomy and contralateral approaches.Use of undersized guidewires may lead to insufficient support of the device which can compromise stent delivery.This stent was used with a.014 wire.
 
Event or Problem Description
It was further reported that the deployment thumbwheel became increasingly more difficult as deployment was initiated.The pull grip eventually disconnected due to the high forces required.The patient's calcification and disease was descried as very difficult and 'hard as a rock'.The physician did assess this as being a factor in the procedure.The patient was confirmed to be fine.No complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA¿
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7771926
Report Number2134265-2018-07262
Device Sequence Number11250390
Product Code NIP
UDI-Device Identifier08714729873952
UDI-Public08714729873952
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/13/2021
Device Model NumberH74939293051530
Device Catalogue Number39293-05153
Device Lot Number0022112677
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2018
Initial Date Received by Manufacturer 07/19/2018
Supplement Date Received by Manufacturer07/26/2018
09/05/2018
Initial Report FDA Received Date08/10/2018
Supplement Report FDA Received Date08/17/2018
09/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-