An investigation was performed for the reported customer complaint: ¿customer states that the power cord plug is broken.Upon triage it was noted that the live pin on the plug is broken.¿ a review of the device history record (dhr) for lot no.V1121056sx indicated the product was released meeting all quality standards.All dhrs are reviewed for accuracy prior to product release.In-process procedures are also in place to prevent nonconforming product in the manufacturing process.This ensures components and finished products meet all quality inspection standards.These controls include, but are not limited to: material verification/certification processes, dimensional specifications, statistical samplings, periodic audits, process inspections, machine maintenance/operation and personnel training and certification.The unit was returned for preventative maintenance and evaluated for the reported customer complaint.Visual inspection observed the bed hook and the plug of the power cord were broken.The bed hook and power cord were replaced.After component replacement, the device was fully tested for functionality and met manufacturer's guidelines.In addition, electrical safety testing was performed pursuant to iec62353 with passing results.The reported customer complaint is confirmed.The root cause is determined to be customer misuse.This complaint will be used for tr ending purposes.If information is provided in the future, a supplemental report will be issued.
|