• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977A260
Medical Device Problem Codes Difficult to Insert (1316); Positioning Problem (3009)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) regarding a patient with an implantable neurostimulator (ins) via a manufacturer representative.It was reported that the stylet did not stay seated in the lead steering hub, making steering the lead difficult and causing delays in the procedure.The procedure needed to be stopped multiple times to reseat the lead.The manufacturer representative indicated that design of the lead is the problem.The lead was clamped into place so that it would stay seated and the lead could be steering.No symptoms or complications were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received from a manufacturer representative (rep) on 2018-aug-10.The rep indicated that the design of the lead hub is the product problem.The patient¿s weight was not available due to hipaa.The device would not be returned for analysis because it was used in the case.This information was confirmed with the healthcare professional (hcp).No further complications were reported/are anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VECTRIS SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7772321
Report Number2649622-2018-13096
Device Sequence Number1966357
Product Code LGW
UDI-Device Identifier00643169109391
UDI-Public00643169109391
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/05/2022
Device Model Number977A260
Device Catalogue Number977A260
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer08/10/2018
Initial Report FDA Received Date08/10/2018
Supplement Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
-
-