• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PDS+ VIO 27IN 2-0 S/A CT-2; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PDS+ VIO 27IN 2-0 S/A CT-2; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Catalog Number PDP333H
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent a gall bladder procedure on (b)(6) 2018 and suture was used.The suture snapped away from the needle when closing the umbilicus.A different suture was used to complete the procedure.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 12/16/2018.Evaluation: a top foil, a paper lid, a winding former and a needle/suture piece returned for analysis.During the visual inspection of sample, the swage and attachment area were as expected.The barrel hole was examined under magnification for suture remnant and still had attached a suture piece to barrel hole.No needle pull off could be observed.The suture piece was examined and the end present elongation and the end appeared to be damaged (dent) by the use of a surgical instrument.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition, no needle suture pull off was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDS+ VIO 27IN 2-0 S/A CT-2
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7773585
Report Number2210968-2018-75070
Device Sequence Number1012378
Product Code NEW
UDI-Device Identifier10705031048065
UDI-Public10705031048065
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K061037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Catalogue NumberPDP333H
Device Lot NumberLPZ330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Initial Date Received by Manufacturer 07/19/2018
Supplement Date Received by Manufacturer11/21/2018
Initial Report FDA Received Date08/10/2018
Supplement Report FDA Received Date12/16/2018
Patient Sequence Number1
-
-