AV-TEMECULA-CT HI-TORQUE PILOT GUIDE WIRE Back to Search Results
Catalog Number 1010480-H
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.The reported difficult to position and difficult to remove were able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that inadvertent mishandling and/or interaction with other devices resulted in the noted bent core; thus resulting in the reported difficult to position during advancement and the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat chronic total occlusion (cto) in the left anterior descending (lad) coronary artery.At the end of procedure, a new ht pilot 50 guide wire was advanced to the lesion without any issues noted.A non-abbott intravascular ultrasound (ivus) device made one successful pass over the ht pilot 50 guide wire.During the second pass, resistance was felt between the ht pilot 50 guide wire and the ivus during advancement.The ivus had to be pulled using force to be removed from the guide wire during withdrawal.Both devices were removed individually from the patient anatomy.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE PILOT GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7773949
Report Number2024168-2018-06181
Device Sequence Number1012849
Product Code DQX
UDI-Device Identifier08717648073854
UDI-Public08717648073854
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K060449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue Number1010480-H
Device Lot Number8052971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2018
Initial Report FDA Received Date08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-