COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Model Number G52272
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Materials coordinator.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the advance 35 lp low profile balloon catheter ruptured in an arteriovenous fistula procedure.Upon reaching the target site in the arm, the device was inflated to 17 atmospheres of pressure, with a "50/50 mix" of contrast and saline, when it ruptured.The rupture was circumferential, making it difficult to retrieve from the sheath, as stated by the initial reporter.There was no angulation, although calcification was noted.The device was removed, and the procedure was completed.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Correction: operator of device: this was inadvertently marked 'ni' when our complaint form did indicate a physician using the device.This field has been updated to 'health professional'.Investigation/evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), manufactures instructions, quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that the balloon had ruptured circumferentially; however, it was also noted that the balloon rupture longitudinally as well.Shaft damage was present and attributed to the device being used and having trouble to remove the device.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed three nonconformances that were found during the manufacturing process on three devices.One of these is related to balloon leakage from the distal bond of the balloon.Neither this nonconformance, nor the others are believed to be associated with the failure mode as all devices were scrapped and did not present with circumferential ruptures.It should be noted there were no other reported complaints for this lot number.Furthermore, the during the device failure analysis, due to the balloon material showing evidence of both a longitudinal and circumferential rupture, it could not be functionally tested.However, both marker bands were present, but were not secure and were moving freely.Moreover, the catheter shaft was elongated between the two balloon bonds.A compression and kink of the catheter shaft was found proximal to the balloon.There appears to be no lost material from the balloon catheter.Confirming that no section(s) of the device remained inside the patient.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7774025
Report Number1820334-2018-02427
Device Sequence Number16397055
Product Code DQY
UDI-Device Identifier10827002522722
UDI-Public(01)10827002522722(17)200406(10)7808619
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2020
Device Model NumberG52272
Device Catalogue NumberPTA5-35-135-7-4.0
Device Lot Number7808619
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer10/17/2018
Initial Report FDA Received Date08/10/2018
Supplement Report FDA Received Date10/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-