• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN DOMINICAN REPUBLIC INC. SAFEDAY¿; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN DOMINICAN REPUBLIC INC. SAFEDAY¿; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number 352894
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up report will be submitted when the results of the investigation become available.
 
Event or Problem Description
As reported by the user facility: as the nurse was administrating the adriamycin iv push, the ll adapter stripped off the y-site of the safeday set, causing the adriamycin to leak.A chemo kit was used to clean up the spill.No injuries were reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample without packaging was returned for evaluation.The set was spiked into a baxter 100ml pab of sodium chloride and a loose teva adapter was also returned.The returned iv set was leakage tested, per specification, with passing results.Both safeday injection sites were accessed with a syringe and the set was tested a second time with passing results.Therefore, the reported defect was not able to be replicated or confirmed.Review of the discrepancy management system (dsms) database performed for the reported lot number revealed no abnormalities or non-conformances noted during in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDAY¿
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
MDR Report Key7774086
Report Number9614279-2018-00130
Device Sequence Number11110689
Product Code FPA
UDI-Device Identifier04046964181894
UDI-Public04046964181894
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K955585
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number352894
Device Lot Number0061619866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2018
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer07/18/2018
Initial Report FDA Received Date08/10/2018
Supplement Report FDA Received Date09/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-