Model Number 1885061HS
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the returned device indicated that the inner shaft broke 0.704¿ from the distal face of the inner hub which would have resulted in the reported malfunction.There were striations around the outside diameter of the break point indicating metal on metal contact during use.The break point corresponds to the proximal end of the outer tube in the front hub.The diamond tip of the bur was compacted with biological contaminants.When viewed under magnification, there was deformation and indentations of the front hub locking areas consistent with aggressive use.There were no bends or signs of concentricity issues.There was an out of specification condition identified as it relates to the complaint.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional reported via a manufacturing representative (rep) that two new burs were broken during an operation.It was noted that the doctor dealt with the issue in time and there was no damage to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7774251
Report Number1045254-2018-00361
Device Sequence Number1012689
Product Code EQJ
UDI-Device Identifier00681490693066
UDI-Public00681490693066
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2026
Device Model Number1885061HS
Device Catalogue Number1885061HS
Device Lot Number0214882498
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2018
Initial Report FDA Received Date08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-