• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number 63430
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a zero tip stone retrieval basket was used in the ureter for a laser ureteroscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, a zero tip the basket was used in an attempt to remove a 4mm sized stone.However, the stone was too big to be removed from the ureter.The wire of the basket was cut to release the stone but failed to do so.A semi-rigid scope was then inserted into the ureter and the use of fiber was used to break up the stone, and complete the case.There were no serious injury or adverse patient effects reported as a result of this event.The patient's outcome at the conclusion of the procedure was reported to be no serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZERO TIP
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
100 boston scientific way
marlborough MA 01752
MDR Report Key7774313
Report Number3005099803-2018-60003
Device Sequence Number16535485
Product Code FFL
UDI-Device Identifier08714729342007
UDI-Public08714729342007
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/07/2021
Device Model Number63430
Device Catalogue Number63430
Device Lot Number0022225195
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2018
Initial Report FDA Received Date08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-