BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493926220220
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/19/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
It was reported that stent damage occurred.The target lesion was located in the severely tortuous and non calcified circumflex artery.A 2.25 x 20mm synergy¿ drug-eluting stent was advanced but was unable to reach the lesion.Eventually, it was noted that the stent was lifted.No patient complications were reported.
 
Manufacturer Narrative
Device avail.For eval, returned to mfr.On, device returned to mfr., device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr: synergy ous mr 2.25 x 20mm stent delivery system (sds) was returned for analysis.A visual and microscopic examination of the crimped stent identified stent damage.Damage was noted to both the proximal and distal ends of the stent; the 2 most proximal stent strut rows were damaged and pulled in a distal direction.The distal end of the stent was stretched and pulled distally over the distal markerband.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.A visual and tactile examination of shaft polymer extrusion revealed no issues.A visual and microscopic examination found damage to the tip.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was further reported that the procedure was discontinued without deploying any stents and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7774420
MDR Text Key116866796
Report Number2134265-2018-07161
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/03/2019
Device Model NumberH7493926220220
Device Catalogue Number39262-2022
Device Lot Number0021586031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2018
Initial Date Manufacturer Received 07/19/2018
Initial Date FDA Received08/10/2018
Supplement Dates Manufacturer Received08/12/2018
Supplement Dates FDA Received09/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-