| Model Number |
C407699 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Occlusion (1984); Stenosis (2263); Vascular Dissection (3160)
|
| Date of Event |
07/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
After a successfull portico implant, the ultimum 19f was removed.An angio control has showed a complete occlusion of the left cfa.A tea and an artery reconstruction with a biocompatible tissue was performed.An angio post intervention showed a left cfa stenosis: a pta was performed with a 7 mm balloon with good final result.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received that this event was not related to the ultimum introducer and is therefore not reportable to the ultimum introducer.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a 27 mm portico valve was implanted with a 19 fr portico delivery system.After a successful portico implant, the ultimum ev hemostasis introducer was removed.An angio control showed complete occlusion of the left cfa.A tea and an artery reconstruction with a biocompatible tissue was performed.An angio post intervention showed a left cfa stenosis and a pta was performed with a 7 mm balloon with good final result.The patient is reported to be recovering.The operators and the vascular surgeon believes the closure systems (2 proglide) has moved a huge calcification, next to the femoral bifurcation, that has occluded the cfa.On (b)(6), the physician indicated the cause of the occlusion was a dissection that was created during use of the portico delivery system or ultimum ev hemostasis introducer.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of a vessel occlusion and dissection was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|