• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VERSALOK PEEK WITH ORTHOCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VERSALOK PEEK WITH ORTHOCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 210818
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/13/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and evaluated.Visual inspection of the device revealed one of the threader wire loops on the threader tab was broken, confirming this complaint.Visual inspection of the returned versalok anchor and inserter revealed no anomalies.It was reported that the wire broke on the suture; a potential root cause for the device failure is more than two strands being threaded through the wire loop, which can become jammed in the eyelet of the anchor resulting in the wire break.Other than this possibility, a definitive root cause for this device failure cannot be determined.Review of the device history record indicated that this batch of product was processed without incident; therefore there is no evidence of manufacturing anomalies on the records reviewed.Review of the depuy synthes mitek complaints system revealed no other complaints of any type for this lot of devices released to distribution.At this time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family to monitor the extent to which this complaint is observed in the field.(b)(4).This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the sales rep via phone call that their versalok peek with orthocord failed at a rotator cuff repair for their customer.The wire broke on the suture prior to using it on the patient once removed from packaging.The procedure was completed using another device.There was no adverse patient consequence nor time delays reported.The device will be returned for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSALOK PEEK WITH ORTHOCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7775618
Report Number1221934-2018-53076
Device Sequence Number1319939
Product Code MAI
UDI-Device Identifier10886705001323
UDI-Public10886705001323
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2018
Device Catalogue Number210818
Device Lot Number3850491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2017
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/13/2016
Initial Report FDA Received Date08/11/2018
Patient Sequence Number1
-
-