| Catalog Number |
K942H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.Did any piece of the needle fall inside of the patient? if so, was the needle piece removed and how? 2.Was there any medical or surgical intervention?.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2018 and suture was used.During the procedure, the needle pulled off from the thread, tearing the mucous membrane.There were no adverse patient consequences reported.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Pc-(b)(4) date sent to the fda: (b)(4) 2018.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).Date sent to fda: (b)(4) 2018.Additional information was requested and the following was obtained: was there any medical or surgical intervention? none intervention on the patient has been necessary.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 12/05/2018.Investigation summary: it was reported needle pull off.A labeled winding former, a needle/suture piece, a suture piece and six unopened samples of product code k942, lot # lgp500 were returned for analysis.During the visual inspection of needle/suture piece and a suture piece, the swage and attachment area were as expected.However, chipping defect was noted; since, suture is swaged excessively at the end of barrel causing some degree of suture damaged.The suture pieces were examined and the ends were cut that appear to be by the use of a surgical instrument.Functional test was performed on the sample and the pull force met the requirements.In the visual inspection of six unopened samples, no defects were found on the packages.The samples were opened and the swage and attachment area were as expected.However, chipping defect was noted; since, suture is swaged excessively at the end of a channel/barrel causing some degree of suture damage.The suture was dispensed without problems and examined along of the strand and no defects were observed.Functional test was performed on the samples and the pull force met the requirements.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the one needle/suture piece and six representatives samples, the assignable cause of the needle pull off was caused by chipping defect.
|
| |
|