MEDTRONIC IRELAND RESOLUTE INTEGRITY RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RSINT27522UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Patient Problem/Medical Problem (2688)
Date of Event 07/27/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During the procedure, an attempt was made to use one resolute integrity rx coronary drug eluting stent to treat a moderately tortuous and none calcified lesion located on the proximal lad artery exhibiting 99% stenosis.There were no abnormalities in relation to the anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop/tray.The device was inspected with no issues noted.The lesion was pre-dilated.The device did not pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was not used during delivery.It was reported that stent dislodgement occurred during delivery to the lesion.The dislodged stent migrated to the diagonal branch and was crushed against the vessel wall using a similar sized stent.The procedure was completed using a third stent prior to crushing the dislodged stent.Patient status post-procedure is alive with no injury.
 
Additional Manufacturer Narrative
Correction: if information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE INTEGRITY RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7777078
Report Number9612164-2018-02022
Device Sequence Number1319589
Product Code NIQ
UDI-Device Identifier00643169037458
UDI-Public00643169037458
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P110013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/14/2019
Device Model NumberRSINT27522UX
Device Lot Number0008585878
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer01/11/2019
Initial Report FDA Received Date08/13/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient Weight72
-
-