A product recall letter was issued to all architect customers who have received architect creatine kinase, list number 07d63-21, lot number 99632un18.The letter informs the customer of the issue regarding a stability issue that may lead to quality control results shifting low out of range and/or error code 1054 "unable to calculate result, reaction check failure" for quality control and patient samples.The letter instructs the customer to discontinue use of the suspect lots and destroy any remaining inventory.The cause of the stability failure has been traced to a contamination of the bulk material.
|
The customer observed biorad qc shifting low out of range and error code 1054 unable to calculate result, reaction check failure for patient and control results.The issue occurred while using the architect creatine kinase assay (list number 7d63, lot number 99632un18).No impact to patient management was reported.
|