(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Visual and functional inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other incidents.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.It is likely that the reported resistance and balloon rupture is the result of interaction with anatomy which was described as heavily stenosed with heavy calcification.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|