Model Number 1000
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio determined that the cause of the reported issue was due to the membrane switch on the device's control panel was shorted.The customer received a replacement device.
 
Event or Problem Description
The customer contacted physio-control to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the device would not power on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 1000 DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7778246
Report Number0003015876-2018-01301
Device Sequence Number1231008
Product Code MKJ
UDI-Device Identifier00885074698878
UDI-Public00885074698878
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1000
Device Catalogue Number99425-000023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2018
Initial Report FDA Received Date08/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-