| Model Number |
1000 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio determined that the cause of the reported issue was due to the membrane switch on the device's control panel was shorted.The customer received a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the device would not power on.
|
| |
|