MEDTRONIC XOMED INC. NIM® PROBE; STIMULATOR, NERVE Back to Search Results
Model Number 8225101
Medical Device Problem Code Intermittent Continuity (1121)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The completed analysis confirmed that one opened sample of part number 8225101 from lot number 0215326232.The reported failed was that the probe was not responding, and visually, there was no damage or anomalies in the construction of the device that would have resulted in the reported event.The tip fit securely in the handle, and the measured resistance was within specification; there was no intermittent behavior while manipulating the tip, connector, and wire.The probe was plugged into a nim system (with a patient simulator) using 1.0 ma stimulating current and 100uv threshold; when stimulating, the system consistently returned 1.02 ma and a waveform/event tone over 300uv.There was no evidence of improper manufacturing and no malfunction found as it relates to the complaint therefore manufacturing and supplier issues have been ruled out as likely causes.There was no out of specification condition identified as it relates to the complaint.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Using the stimulating probe, the doctor was not getting readings occasionally; the nim unit was tested with a new, different probe without issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® PROBE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7778384
Report Number1045254-2018-00370
Device Sequence Number19281045
Product Code ETN
UDI-Device Identifier00681490176682
UDI-Public00681490176682
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K934426
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/22/2026
Device Model Number8225101
Device Catalogue Number8225101
Device Lot Number0215326232
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2018
Initial Report FDA Received Date08/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-