| Model Number |
MMT-551NAP |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump has a crack.The customer¿s blood glucose level was 180 mg/dl.Troubleshooting was performed for damage and noticed the reservoir was unable to lock in place.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with broken reservoir tube lip.A test reservoir was installed and did not lock in place due to complete broken reservoir tube lip.
|
| |
|