• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 133PLUS SV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 133PLUS SV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133P-SV-16-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: deflation: patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.The following is a list of potential adverse events that may occur with breast implant surgery.The risks include: implant deflation/leakage, additional surgery, capsular contracture, infection, toxic shock syndrome, necrosis, hematoma, seroma, extrusion, breast pain, changes in nipple sensation, changes in breast sensation, dissatisfaction with cosmetic results (wrinkling, folding, displacement, asymmetry, palpability, visibility, ptosis, sloshing), calcific deposits, irritation/inflammation, delayed wound healing, hypertrophic scarring, breast tissue atrophy/chest wall deformity, difficulty/inability in breast feeding, and inability to adequately visualize breast lesions with mammography.In addition to these potential adverse events, there have been concerns with certain systemic diseases.
 
Event or Problem Description
Health professional reported right side device removal due to "defective implant deflating." the device was removed and replaced.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified opening and fluids.A leak test was performed which found opening on radius of the device.Microscopic analysis was performed which identify a striated edge opening, consist with use of a surgical tool on radius side.
 
Event or Problem Description
Health professional reported right side device removal due to "defective implant deflating." the device was removed and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133PLUS SV TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key7778708
Report Number9617229-2018-05890
Device Sequence Number19281093
Product Code LCJ
UDI-Device Identifier10888628010352
UDI-Public10888628010352
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K143354
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/27/2021
Device Catalogue Number133P-SV-16-T
Device Lot Number2941617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Initial Date Received by Manufacturer 07/17/2018
Supplement Date Received by Manufacturer08/28/2018
Initial Report FDA Received Date08/13/2018
Supplement Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-