• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-4316
Device Problem Component Missing (2306)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/06/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Device evaluation indicated that one of the eyelets of the clik anchor was torn apart with missing silicone material.Sc-1132 (sn:(b)(4)) and sc-2218-70 ((b)(4)) device evaluation indicated that the ipg and leads passed visual, electrical and performance tests performed.
 
Event Description
A report was received that the patient underwent an explant procedure to be eligible for mri.It was found out through the device evaluation that one of the eyelets of the clik anchor was torn apart with missing silicone material and unknown if this was been removed from the patients body.No further information can be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key7779169
MDR Text Key116993255
Report Number3006630150-2018-02316
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729793267
UDI-Public08714729793267
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/20/2017
Device Model NumberSC-4316
Device Lot Number17900199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/23/2018
Initial Date FDA Received08/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age39 YR
-
-