MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH LOCKING COND PL 1.5 SHAFT 6HO HE 2HO SST; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH LOCKING COND PL 1.5 SHAFT 6HO HE 2HO SST; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.114.014
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device used in a veterinary case - no patient information will be reported.Device malfunctioned intra-operatively and was not implanted / explanted.Date returned to manufacturer.A review of the device history records has been requested.Subject device has been received and is currently in the evaluation process.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This report is for a veterinary case.There was no human patient involvement.Device report from synthes europe reports an event in (b)(6) as follows: it was reported that patient underwent an osteosythesis for distal radius fractures on (b)(6) 2018.During the procedure, surgeon attached two (2) bending pins to a plate, then tried to contour the plate using the bending pins.However, the bending pins came off the plate while bending.Surgeon attempted to use two (2) other plates but obtained the same results.It was also reported that during the same procedure, a locking screw would not lock to the plates.Surgeon was not sure if the issue was due to the bending pins or the screw hole on the plates.Surgery was completed with a delay of approximately 30 minutes and no reported consequence to the patient.This report is for one (1).This is report 1 of 5 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record (dhr) review was conducted.Part mfg date: 05-nov-2015, part exp.Date: n/a, manufacturing location: (b)(4).Lot number: 9935795, part number: 02.114.014.Dhr record review: a review of the device history record (dhr) revealed no non-conformances.The dhr shows lot#: 9935795 of 1.5mm lcp condylar plate 6 hole shaft was processed through the normal manufacturing and inspection operations with no non-conformances or rework noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material dhr revealed this lot met all specifications with no non-conformances or rework noted.A product investigation was conducted: visual inspection: three plates were involved in the event and were returned.Per design, the plates are not etched with lot numbers, therefore, this investigation contains results for all three plates.Visual inspection of the returned plates revealed stripping of the thread-starts of all holes for all plates.One plate¿s plate head was received bent: the two outer edges of the head¿s tangs were bent downward, as if they were being contoured to the bone.Another plate was received with a slight bend longitudinally.The third plate was received with no visible bend.The complaint condition specifies that the issue of ¿unable to assemble¿ occurred during plate bending, which is an intended use of the plate described in technique guides.Therefore, no issue has been identified regarding the plates¿ bending.Functional test: functional testing with the two returned bending pins (03.114.003/lot: h299829) and two locking screws (02.214.012/lots: 7086832 and 7815462) was performed.Difficulty engaging the threads of the bending pins and locking screws was experienced with all plates in all holes.The received condition for all three plates agrees with the complaint description, the complaint was replicated with the returned devices.Dimensional inspection: dimensional inspection was not performed due to damage.Document/specification review: device history record(s) for lots: 9935795 and h322156 showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.The lot number was unknown for the third plate, therefore review of the device history record could not be performed.The relevant drawings were reviewed: (current at the time of manufacture and investigation for the two plates with known lot numbers; as the lot number was unknown for the third plate, the drawing used at the time of manufacturer could not be determined).During the investigation no product design issues were observed that may have contributed to the complaint condition.Conclusion: the complaint condition was confirmed with investigation.There were no issues during the manufacture of this product that would contribute to this complaint condition.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.Dimensional inspection of the returned device revealed no non-conformances of the device¿s geometric specifications.While no definitive root cause could be determined, it is possible that off axis thread-engagement between the bending pins and plates stripped the plates¿ thread-starts.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING COND PL 1.5 SHAFT 6HO HE 2HO SST
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7779194
Report Number8030965-2018-55696
Device Sequence Number19140156
Product Code HRS
UDI-Device Identifier10886982034922
UDI-Public(01)10886982034922
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K090047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.114.014
Device Catalogue Number02.114.014
Device Lot Number9935795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2018
Initial Date Received by Manufacturer 07/29/2018
Supplement Date Received by Manufacturer09/25/2018
Initial Report FDA Received Date08/13/2018
Supplement Report FDA Received Date10/08/2018
Patient Sequence Number1
-
-