• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OHIO MEDICAL LLC CARE-E-VAC 3; PORTABLE SUCTON DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OHIO MEDICAL LLC CARE-E-VAC 3; PORTABLE SUCTON DEVICE Back to Search Results
Model Number 758000
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Field correction 1419185-7/24/18-001-c has been initiated for correction of other potentially affected devices.
 
Event or Problem Description
Complaint was received on (b)(6) 2018 stating "after final visual inspection of the care e vacs a total of three units all have discoloration to the casing, and smell as if the unit has been shorted out.Upon return to ohio medical on july 19, 2018 and repair center evaluation on july 24, 2018 it was noted "due to extent of damage, unit is beyond economical repair".Further evaluation was performed by engineering which determined that a short occurred when battery terminals suspectedly came into contact with pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARE-E-VAC 3
Common Device Name
PORTABLE SUCTON DEVICE
Manufacturer (Section D)
OHIO MEDICAL LLC
1111 lakeside dr
gurnee IL 60031
MDR Report Key7779345
Report Number1419185-2018-00008
Device Sequence Number2447054
Product Code BTA
UDI-Device Identifier08688250002436
UDI-Public(01)08688250002436
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K062610
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Remedial Action Other,Recall
Type of Report Initial
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number758000
Device Catalogue Number758000
Device Lot NumberCEV318020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/12/2018
Initial Report FDA Received Date08/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-