| Catalog Number |
AB35W08040135 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician was attempting to use an evercross balloon dilation catheter to treat a moderately calcified mid right common iliac artery.The vessel had little tortuosity and the lesion had resulted in 90% stenosis.The device was prepped with no issues identified.There was no resistance encountered advancing the device and it did not pass through a previously deployed stent.During the first balloon inflation at 7atm the balloon burst.The procedure used a non-mdt inflation device.There were no balloon fragments.All of device was withdrawn from the patient without intervention.The artery was stented with a 10x40 protege, and post dilated with a competitive balloon to complete the procedure.The use of the protege was planned.No patient injury was reported.
|
| |
|