| Model Number |
G22170 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
This complaint is being reported under the fda mdr malfunction precedence of "stent kinked/ bent".As reported to customer relations: "devices appeared kinked in the pigtail.Same place in each stent.Dm confirmed some sort of kink.Occurred when preparing devices, so could have been user-related.It's on the proximal pigtail.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Importer site establishment registration number: (b)(4).Additional information received: please indicate any other endoscopic accessories (if any) that came into contact with the stent or introduction system during the procedure.-- hydra jag wire 260.
|
| |
|
Event or Problem Description
|
|
This follow up is being submitted as additional information was received on the 18-sep-18 regarding the wire guide used.This complaint is being reported under the fda mdr malfunction precedence of "stent kinked/ bent." as reported to customer relations: "devices appeared kinked in the pigtail.Same place in each stent.Dm confirmed some sort of kink.Occurred when preparing devices, so could have been user-related.It's on the proximal pigtail.
|
| |
|
Event or Problem Description
|
|
This complaint is being reported under the fda mdr malfunction precedence of "stent kinked/ bent".As reported to customer relations: "devices appeared kinked in the pigtail.Same place in each stent.Dm confirmed some sort of kink.Occurred when preparing devices, so could have been user-related.It's on the proximal pigtail.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(6).(b)(4).Importer site establishment registration number: (b)(4).Additional information received: where was the device was stored prior to use.Storage cabinet were previous procedures i.E.Sphincterotomy etc.Carried out prior to placing the device over the particular wire guide? if yes please indicate the procedure performed.Ercp was the wire guide inspected for damage prior to use?yes was the device at the centre of the complaint inspected for damage prior to use?the first stent was not, but the second one was did the patient involved exhibit altered anatomy or tortuous anatomy?no.If not with the device in question, how was the procedure finished?chose another stent had a sphincterotomy been performed prior to this occurrence?no.What is the endoscope manufacturer, the model number and working channel size that was used for the procedure?olympus tjf-180.Does your medical facility have a service/maintenance schedule associated with its endoscopes?yes.Please indicate the location in the body where the stent device was to be placed.I.E.Biliary duct, pancreatic duct, other.Biliary duct.Was resistance encountered when advancing the wire guide to the target location? was the stricture dilated prior to placing the device? if so, please indicate what device(s) were used.Was resistance encountered when advancing the introduction system in place to the target location? was resistance encountered when advancing the stent through the obstructed area? (mark n/a if device was not used on any patient) after placement, was stent position verified? if yes, please describe how.Did any section of the device detach inside the endoscope or patient? please indicate any other endoscopic accessories (if any) that came into contact with the stent or introduction system during the procedure.Lab evaluation: the device zso-7-9 of lot # c1379490 was returned to cirl for evaluation.During the lab evaluation, the stent was found to be kinked at first porthole (non tapered) end and fifth porthole (tapered) end.Wire guide passed when loaded from non tapered end.Wire guide stopped at fifth porthole when loaded from tapered end.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However, a possible cause of this complaint is that the physician may have kinked the device when removing it from the packaging.The complaint is confirmed as the failure was verified in the laboratory and the stent was found to be kinked.Documents review: a review of the manufacturing records for the device of lot # c1379490 did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1379490; upon review of complaints this failure mode has not occurred previously with this lot # c1379490.Prd/fqc/pkg review: prior to distribution, all devices are subjected to 100% inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.There is also a visual inspection of the product and packaging at packaging, packaging qc, and post sterile qc.Ifu review: it may be noted that according to the instructions for use, ifu0045-6, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.Summary: the complaint is confirmed as the failure was verified in the laboratory and the stent was found to be kinked.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|