• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE Back to Search Results
Model Number N/A
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 07/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2016018 b.Braun medical, inc.(importer) is submitting this report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun melsungen ag internal report # (b)(4).We received no sample and not picture.Because of this, further investigation of the complaint is not possible.Device history record (dhr): reviewed the device history record and no abnormalities found during in process and final control inspection.
 
Event or Problem Description
As reported by the user facility (translation of user facility information by bbm sales organization in spain: safety device of introcan safety g18 1,3x32 mm-pur doesn´t work propertely and a nurse has a puntion in operating room.
 
Additional Manufacturer Narrative
Exemption number e2016018.B.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun medical inc.Internal report number (b)(4).This follow up report is being filed to correct the branding, product code, and 510(k) information that was entered on the initial mdr in sections d1, d2, and g5.This report was filed for item number, 4251687-01, which is not sold in the united states, however a similar item number, 4251687-02, is sold in the united states by b.Braun medical, inc.The 510(k) has been updated to reflect a 510(k) that item 4251687-02 has been cleared under.Note: b.Braun is aware that the listed exemption has been withdrawn, however since the initial mdr report was filed under this exemption, the follow up is also being filed under the exemption.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key7780742
Report Number9610825-2018-00163
Device Sequence Number16743707
Product Code FOZ
UDI-Device Identifier04046964381997
UDI-Public(01)04046964381997(17)230201(10)18B08G8271
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/02/2019
Report Date (Section F) 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2023
Device Model NumberN/A
Device Catalogue Number4251687-01
Device Lot Number18B08G8271
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/14/2018
Device Age6 MO
Event Location Hospital
Date Report to Manufacturer07/02/2019
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer07/20/2018
Initial Report FDA Received Date08/14/2018
Supplement Report FDA Received Date07/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA07/02/2019
-
-