DEPUY ORTHOPAEDICS, INC. 1818910 FEM/TIB EXTRAC 11X7 3/4 IN; KNEE INSTRUMENT : EXTRACTION INSTRUMENTS Back to Search Results
Catalog Number 242102000
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Fem/tib extractor handle cracked and the tip of the instrument broke off while removing a femoral component.All pieces were retrieved and there was no additional time added to the surgery.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirms the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEM/TIB EXTRAC 11X7 3/4 IN
Common Device Name
KNEE INSTRUMENT : EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7781320
Report Number1818910-2018-66600
Device Sequence Number13747062
Product Code HWB
UDI-Device Identifier10603295122722
UDI-Public10603295122722
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242102000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2018
Initial Date Received by Manufacturer 07/19/2018
Supplement Date Received by Manufacturer08/17/2018
Initial Report FDA Received Date08/14/2018
Supplement Report FDA Received Date08/20/2018
Patient Sequence Number1
-
-