| Model Number |
LVZRXT180J |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during the implant procedure, the guidewire broke.The guidewire was replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the guidewire was returned and analyzed.The analysis indicated that the guidewire was unraveled.The guidewire was damaged.Visual analysis of the lead indicated damage during use.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|