• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIAADAPTXL
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic sleeve gastrectomy procedure, for the 2nd firing on the stomach, the device did not advance and unable to fire.The handle was replaced with a manual handle to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual and functional evaluation of the handle noted no abnormalities.Evaluation of the eeprom noted that the code "drive_motor_excessive_load_mode" was present.This was indicative of a thick tissue application and could confirm the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.The file was concluded to be the device not being used as intended.Replication of the drive_motor_excessive_load_mode error codes may occur of the idrive ultra powered handle stops firing before the lower clamp reaches the distal end, which is likely due to the firing force reaching the upper design limit of the endo gia reload.This may occur under the following conditions.Application over tissue that is beyond the recommended thickness range.Application with an obstacle incorporated in the jaws.In either scenario, staples may not form properly and tissue may not be fully transected.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7781523
Report Number1219930-2018-04467
Device Sequence Number16523149
Product Code GDW
UDI-Device Identifier10884521201989
UDI-Public10884521201989
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K133762
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEGIAADAPTXL
Device Catalogue NumberEGIAADAPTXL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer01/14/2019
Initial Report FDA Received Date08/14/2018
Supplement Report FDA Received Date06/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-