| Model Number |
EGIAADAPTXL |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic sleeve gastrectomy procedure, for the 2nd firing on the stomach, the device did not advance and unable to fire.The handle was replaced with a manual handle to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual and functional evaluation of the handle noted no abnormalities.Evaluation of the eeprom noted that the code "drive_motor_excessive_load_mode" was present.This was indicative of a thick tissue application and could confirm the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.The file was concluded to be the device not being used as intended.Replication of the drive_motor_excessive_load_mode error codes may occur of the idrive ultra powered handle stops firing before the lower clamp reaches the distal end, which is likely due to the firing force reaching the upper design limit of the endo gia reload.This may occur under the following conditions.Application over tissue that is beyond the recommended thickness range.Application with an obstacle incorporated in the jaws.In either scenario, staples may not form properly and tissue may not be fully transected.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|