• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLINICAL INNOVATIONS, LLC GOLD TRACE FETAL SCALP ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLINICAL INNOVATIONS, LLC GOLD TRACE FETAL SCALP ELECTRODE Back to Search Results
Model Number CNS000004
Health Effect - Clinical Codes Abscess (1690); Purulent Discharge (1812); Staphylococcus Aureus (2058); Swelling (2091); Device Embedded In Tissue or Plaque (3165)
Date of Event 01/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It is very important for the fse to be applied correctly to have the best results of monitoring and removal at the end of delivery.It is imperative that the fse not be rotated more than 1 ½ turn when applying it to the baby's scalp.If the electrode is excessively torqued during (e.G.Rotated > 1 ½ turns clockwise) application to the presenting part or if the electrode was rotated in the incorrect direction when removing from the scalp.The instructions for removal of the fse in the ifu reads as follows: "remove spiral needle by grasping electrode wires as close to fetal presenting part as possible and twist counter-clockwise until free from presenting part".When removing the fse, it is very important to make sure that the clinicians are twisting it counter-clockwise as close to the presenting part, as possible.If it was inserted correctly, it should not take more than a 1 ½ turn to remove it.Education on proper insertion and removal along with reviewing of the ifu's are very important for any clinician using this product.
 
Event or Problem Description
Information from the parents: a scalp electrode was applied on the scalp of my child.Because of bradycardia during contractions.Need to monitoring risk of fetal hypoxia.The (b)(6) 2018 (d12), a sudden tumefaction of 2cm appeared on the occipital part with, at the summit, a small crust corresponding to the electrode positioning.The (b)(6) 2018 (d16), the tumefaction is bigger (2,5 cm) with a purulent flow.Diagnosis: under skin abscess with staphylococcus inoculated by scalp electrode monitoring per partum.Broken fetal scalp electrode under the skin caused tumefaction with a purulent flow.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOLD TRACE FETAL SCALP ELECTRODE
Common Device Name
FETAL SCALP ELECTRODE
Manufacturer (Section D)
CLINICAL INNOVATIONS, LLC
747 w 4170 s
murray UT 84123
MDR Report Key7781977
Report Number1722684-2018-00010
Device Sequence Number11278050
Product Code HGP
UDI-Device Identifier00814247020505
UDI-Public00814247020505
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K030691
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCNS000004
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2018
Initial Report FDA Received Date08/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age1 DA
-
-