BARD ACCESS SYSTEMS PONSKY NBR GASTROSTOMY TUBE 20F; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 000702
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that damage at the distal end of the obturator was found at the time of opening the package.There was no reported patient involvement.
 
Event or Problem Description
It was reported that damage at the distal end of the obturator was found at the time of opening the package.There was no reported patient involvement.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was performed.The lot met all release criteria.Investigation summary: one ponsky nbr 20f non-balloon replacement obturator was returned for evaluation.Visual and microscopic evaluations were performed.The investigation is confirmed for damaged/defective accessory , as it was noted that distal and proximal ends of obturator showed excess material.The definitive root cause was determined to be supplier related where the flash from the molded parts was not removed after processing.Labeling review:a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PONSKY NBR GASTROSTOMY TUBE 20F
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7782267
Report Number3006260740-2018-02058
Device Sequence Number1793594
Product Code KNT
UDI-Device Identifier00801741036293
UDI-Public(01)00801741036293
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103109
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000702
Device Catalogue Number000702
Device Lot NumberREBZ0337
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2018
Initial Date Received by Manufacturer 07/24/2018
Supplement Date Received by Manufacturer05/16/2019
Initial Report FDA Received Date08/14/2018
Supplement Report FDA Received Date05/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-