| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report # (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58 - user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip (b)(4).Note: manufacturer contacted customer on 8/6/2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high and erratic blood glucose test results.The customer is concerned with test results obtained of 115, 163, 118, 213 and 196 mg/dl.The customer was using correct testing technique.The customer's expected fasting blood glucose test result range is 100 - 120 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer was unable to perform a back to back blood test during the call on (b)(6) "2019".The product is stored according to specification in the bedroom.Test strip lot manufacturer's expiration date is 07/29/2019 and open vial date is july 2018.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|