BARD ACCESS SYSTEMS BARD BUTTON REPLACEMENT GASTROSTOMY DEVICE 18F X 1.7CM; TUBE, GASTRO-ENTEROSTOMY Back to Search Results
Model Number 000282
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was not provided, a review of the device history records could not be performed.The device is pending return to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the valve broke after approximately four months of use.There was no reported patient injury.
 
Event or Problem Description
It was reported that the valve broke after approximately four months of use.The device was removed and another manufacturer's device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a complete manufacturing review could not be performed, as the lot number is unknown.Investigation summary: one button replacement 18f 1.7cm gastrostomy non-balloon sample was returned for evaluation.The sample was returned in two segments, with the detached cap plug failure mode being addressed in related complaint mdr number 3006260740-2018-02152.As such, this investigation will not address the detached cap plug.The valve was patent to infusion with no issues noted during testing.Therefore, the investigation is unconfirmed for the reported broken valve.As the sample was received with its cap plug detached, it is possible that damage related to the detachment contributed to the reported event.However, as no other issues could be identified with the valve during evaluation, the definitive root cause could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the reported event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD BUTTON REPLACEMENT GASTROSTOMY DEVICE 18F X 1.7CM
Common Device Name
TUBE, GASTRO-ENTEROSTOMY
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7782340
Report Number3006260740-2018-02065
Device Sequence Number1012436
Product Code KGC
UDI-Device Identifier10801741087858
UDI-Public(01)10801741087858
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K904779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000282
Device Catalogue Number000282
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Initial Date Received by Manufacturer 07/27/2018
Supplement Date Received by Manufacturer02/11/2019
Initial Report FDA Received Date08/14/2018
Supplement Report FDA Received Date02/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-