• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PNEUPAC VENTILATORS VR1 STANDARD; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PNEUPAC VENTILATORS VR1 STANDARD; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Catalog Number 520A1299US
Medical Device Problem Code Tidal Volume Fluctuations (1634)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation in progress.
 
Event or Problem Description
It was reported that the tidal volume reading was high on the pneupac ventilator.No death or serious injury was reported in connection with this incident.
 
Additional Manufacturer Narrative
One pneupac® ventilator was returned in good condition for analysis.Relief valve check was run at the adult setting; device failed.Tidal volume was found to be out of tolerance at child setting.Attempted calibration done of tidal volume but tidal volume was noted to still be out of tolerance.Timing volume assembly part was replaced and the unit passed all functional testing.Based on the evidence, the complaint was confirmed with the root cause being due to normal wear and tear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTILATORS VR1 STANDARD
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7782498
Report Number3012307300-2018-03242
Device Sequence Number1259033
Product Code BTL
UDI-Device Identifier10610586045721
UDI-Public10610586045721
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K051322
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number520A1299US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2018
Supplement Date Received by Manufacturer11/01/2018
Initial Report FDA Received Date08/14/2018
Supplement Report FDA Received Date12/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-