| Model Number |
ZXT225 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Halo (2227); No Code Available (3191)
|
| Date of Event |
04/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: exact date not provided; however reported as 27 weeks post op.(b)(6) 2018 used as an approximate date.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent implant with a model zxt225 17.0d intraocular lens (iol) in her left eye (os).Approximately 27 weeks postop, the subject experienced halos, a lot of glare at night and a lack of near visual acuity.Patient's best corrected distance visual acuity (bcdva) was 20/25.On (b)(6) 2018, lens was exchanged with a 17.5 diopter model zct300 iol and issues have been resolved with the patient's bcdva at 20/15.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on new information received on 8/15/2018, the following section has been updated accordingly: additional information: visual symptoms were first reported on (b)(6) 2018.There was no patient injury reported and no surgical intervention was required.Furthermore, patient is happy post-surgery.Product will not be returned as it is discarded by the surgery center.The intraocular lens (iol) is not returning for evaluation as it is discarded; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|