• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751PNAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751PNAH; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751PNAH
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 07/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they were hospitalized due to high blood glucose on (b)(6) 2018 with blood glucose of 639 mg/dl and then on (b)(6) 2018 with blood glucose of 520 mg/dl.The customer was treated by insulin and intravenous fluids.Troubleshooting was done for high blood glucose and under delivery.Customer was off the insulin pump since (b)(6) 2018.Customer states the drive support cap appears normal.The customer was wearing the insulin pump during the hospitalization.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
The device received with all operating currents within specification and passed functional testing including the rewind, basic occlusion, occlusion, prime or compromised force sensor system, excessive no delivery, self test, unexpected restart error and displacement test.Pump passed the dat test at 0.08735 inches.The device received with cracked case at display window corners, display window, reservoir tube lip and battery tube threads.The device received with minor scratched display window and case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751PNAH
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7783161
Report Number2032227-2018-13123
Device Sequence Number1392157
Product Code OZO
UDI-Device Identifier00643169507326
UDI-Public(01)00643169507326
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-751PNAH
Device Catalogue NumberMMT-751PNAH
Device Lot NumberA3751PNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2018
Initial Date Received by Manufacturer 07/16/2018
Supplement Date Received by Manufacturer09/14/2018
Initial Report FDA Received Date08/14/2018
Supplement Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age32 YR
Patient Weight278
-
-